GV

GeezerVison

The World Wide Web, restored to 800 pixels of glory.

Home|About|Gaming cabinet|Archives|How it works|Starter sites

On this site

  • Sign in
  • ActionsCI/CD and automation
  • CodespacesCloud development environments
  • All Features
  • Documentation
  • Training
  • Certifications
  • Build with Copilot
  • Customize Copilot
  • Integrations
  • Trending repos
  • GitHub Sponsors
  • Developer articles
  • Enterprises

This view

  • Original page
  • New address

GeezerVison is showing github.com — GitHub - MSF3Dprinting/Universal-Mounting-System: Universal Mounting System V1 (UMS V1) is set of mutually compatible attachments, hooks, holders and clamps used to mount various devices and accessories on the wall, beds and other objects. The UMS allows easy attachment, removal and changing of the devices.

BranchesTags
Open more actions menu

Latest commit

 

History

24 Commits
24 Commits

Folders and files

NameName
Last commit message
Last commit date
3D files
3D files
 
 
drawings
drawings
 
 
img
img
 
 
DATASHEET_TEMPLATE.md
DATASHEET_TEMPLATE.md
 
 
LICENSE
LICENSE
 
 
README.md
README.md
 
 

Repository files navigation

Universal Mounting System V1 (UMS V1)

OSHW Certified KE000001

Certified by the Open Source Hardware Association under UID KE000001. The OSHW mark is owned by OSHWA and is not covered by this repository's MIT License, so forks may not display it or the UID; certify your own derivative at https://certification.oshwa.org/.


1. General Information

1.1 Intended Use Case

The Universal Mounting System V1 (UMS V1) is a set of mutually compatible, 3D-printed attachments, hooks, holders and clamps used to mount devices and accessories on walls, hospital beds, trolleys, poles and other objects in the hospital environment.

Each holder consists of a back part and a front part. The back part is attached to the bed, pole, trolley or wall; the front part slides into the back part and holds the specific device it was designed for (e.g. a humidifier bottle, handheld pulse oximeter, or ultrasound transducer). Because all front and back parts share a common sliding interface, they are interchangeable, allowing easy attachment, removal, reconfiguration and cleaning.

Intended situations of use:

  • Low-resource hospital settings where standard mounting accessories are unavailable or unaffordable, making it difficult to arrange and secure medical devices properly. This leads to equipment damage, shortened equipment lifespan, improper use, and IPC issues.
  • Local, on-demand production via FDM 3D printing so facilities can safely mount equipment in a suitable position and location.
  • Multiple-use item; permanent vs. temporary installation is context dependent.

Explicitly out of scope / not intended: The UMS is an equipment-mounting accessory. It is not intended for patient contact, patient support (e.g. as a grab bar or lifting point), drug delivery, diagnosis, or therapy, and it holds only the specific devices listed in the compatibility table (Section 1.4).

1.2 Target Group and Project Classification

Target group (group of impact):

  • Primary beneficiaries: hospital staff in low-resource hospitals and clinics, where properly mounted equipment (humidifiers, pulse oximeters, ultrasound transducers) improves safety, hygiene and device lifespan.
  • Users of the design files: biomedical technicians, hospital maintenance staff, humanitarian-organization logistics/3D-printing teams, and makers producing parts locally on any FDM printer.
  • Gatekeepers: per the repository, deployment requires approval by a Biomed responsible and an IPC responsible before use.

Project classification: Non-invasive, non-sterile, reusable hospital equipment accessory (mounting hardware). It is not an active device, delivers no energy or substances, and has no intended patient contact.

1.2.1 Risk class and region-specific compliance expectations

The following is the project's good-faith self-assessment, not a formal regulatory determination — classification ultimately depends on the claimed intended purpose and must be confirmed with the competent authority or a regulatory professional in each deployment region. Follow national medical-device regulations where they exist; where they do not, follow institutional/organizational approval: Biomed responsible + IPC repsponsible sign-off is mandatory before use per this project's documentation.

1.2.2 State of development

  • State: Product in development
  • CLassification: field-development / humanitarian-use hardware, produced locally under institutional biomed + IPC approval. It is not a cleared, approved, or CE-marked medical device, and no regulatory submission has been made. It is not currently an end-consumer product; it is intended for controlled deployment in facilities with technical oversight.

1.3 Critical Design Thinking Notes

Critical information required for the project to succeed:

  1. Material cannot be substituted casually. Parts are specified in PETG, White/Natural or another light color, chosen deliberately so that soiling and surface imperfections are clearly visible (an IPC feature, not cosmetics). Substituting dark filaments defeats the contamination-visibility design intent; substituting materials with lower chemical resistance may fail under the specified disinfectants.
  2. Do not autoclave. The polymer will deform. Only surface disinfection is permitted (see Section 2.3).
  3. No text or embossed symbols may be added to parts. Recesses and embossing trap contaminants and interfere with surface cleaning. Version/part identification must live in documentation and packaging, not on the part.
  4. Print orientation and geometry are pre-engineered — print as provided.
    • Bottom edges are chamfered for first-layer quality (image).
    • Horizontally printed holes are geometry-compensated for FDM (image).
    • In-built supports are included in the models where needed (image) — slice with no additional supports. Adding slicer supports or re-orienting parts invalidates these optimizations.
  5. Strength-critical slicer settings: 0.2 mm layer height, 4 perimeters* for increased strength. For hook back parts, a brim is recommended for bed adhesion/stability during printing.
  6. Post-processing is mandatory: remove brim, remove stringing/imperfections, and deburr sharp edges before QC.
  7. Aging behavior: the material can become brittle over time — this drives the weekly-inspection requirement on latches and hooks (Section 1.7).
  8. Fit before installation: back/front parts must slide easily but hold firmly; the target device (e.g. humidifier bottle) must be verified to seat stably and remain easily removable before the mount is installed.

1.4 Bill of Materials (BOM)

All structural parts are locally manufactured by FDM 3D printing from the repository's source files (STEP = editable source, STL/3MF = print-ready, GCODE = pre-sliced reference). For locally printed parts, "manufacturer" is the producing site and the "manufacturer part number" is the repository file name at the pinned commit/release.

1.4.1 Printed parts — Back parts (interface to bed / wall / pole / trolley)

# Part Source files (STL/STEP) Compatibility
B1 Double hook extended 3D files/STLs/Back parts/Double hook extended.stl Hospital bed with backrest — Ø32 mm round horizontal tube; trolley — Ø19 mm round horizontal tube
B2 Single hook 3D files/STLs/Back parts/Single hook.stl Ø25 mm round horizontal tube
B3 Hook with support 3D files/STLs/Back parts/Hook with support.stl Hospital bed with backrest, Ø32 mm round horizontal tube; secured with zip tie for more stable positioning
B4 Wall mount with safety lock 3D files/STLs/Back parts/Wall mount with safety lock.stl Any wall surface (fasteners required, see hardware below)
B5 Vertical attachment 20 mm 3D files/STLs/Back parts/Vertical attachment 20mm.stl Vertical tube/pole, 20 mm
B6 Vertical attachment 25 mm 3D files/STLs/Back parts/Vertical attachment 25mm.stl Vertical tube/pole, 25 mm

1.4.2 Printed parts — Front parts (device holders)

# Part Source files (STL/STEP) Held device
F1 Humidifier bottle holder — 49 mm 3D files/STLs/Front parts/Humidifier bottle holder - 49mm.stl Humidifier bottle, Ø49 mm
F2 Humidifier bottle holder — 56 mm 3D files/STLs/Front parts/Humidifier bottle holder - 56mm.stl Humidifier bottle, Ø56 mm
F3 Pulse oximeter holder 3D files/STLs/Front parts/Pulse oximeter holder.stl Masimo handheld pulse oximeter
F4 Ultrasound transducer holder 3D files/STLs/Front parts/Ultrasound transducer holder.stl Abdominal ultrasound transducer

Print-ready bundles: UMS set.3mf (full plate) and reference GCODE UMS set_0.4n_0.2mm_PETG_MK4S_11h37m.bgcode — 0.4 mm nozzle, 0.2 mm layers, PETG, Prusa MK4S, ≈11 h 37 min for the full set.

1.4.3 Materials, hardware and consumables

Item Spec Used for Manufacturer / part no.
PETG filament White, Natural, or light color; 1.75 mm for the reference GCODE All printed parts Any reputable filament manufacturer (commodity item); record brand/batch locally per print run
Bolts/screws M5 bolts or screws Ø4–5 mm; quantity per hole count in the wall-mount model B4 Wall mount with safety lock Generic DIN/ISO hardware;
Zip tie 6 mm width B3 Hook with support (stabilization) Generic
Zip-lock sealable bags Sized to parts Post-production packaging Generic
Disinfectants Surfanios, or bleach 1:10, or ethanol 70 % Cleaning before use and in service Per facility IPC supply
3D printer Any FDM printer Manufacturing Reference machine: Prusa MK4S

1.5 Testing Data / Testing Verification

What a correctly made UMS part does (acceptance criteria):

  1. Visual: smooth surface; no printing artifacts, stringing, or misaligned layer lines.
  2. Dimensional: back and front parts fit into each other before installation; the target device (e.g. humidifier bottle of the matching diameter) fits its holder; the device holds in a stable manner and remains easily removable.
  3. Tolerance: the front part slides easily into the back part yet holds firmly (no rattle, no forcing).
  4. Safety validation: sign-off by the responsible Biomed and IPC advisors before clinical-area use.
  5. In-service verification: weekly checks after installation — no cracks in the printed body, surface cleanliness, and retained mechanical flexibility of latches and hooks.

Status of quantitative test data: the repository currently documents functional/QC verification criteria (above) but does not yet publish quantitative test results (e.g. static load ratings per hook, cyclic insertion/removal counts. Until such data is published, each producing site must perform and record its own verification per the criteria above, and loads must be limited to the specific devices listed in the compatibility tables. Publication of load-test data is tracked as an open issue (Section 1.6.2).

1.6 Risk Assessment / Analysis

Qualitative risk analysis for the declared intended use (repository-declared risk level: 3 on the project's internal scale). Severity/likelihood are estimated qualitatively pending quantitative test data.

# Hazard / failure point Foreseeable harm Est. likelihood Est. severity Mitigations (designed-in / procedural)
R1 Mechanical failure of hook/holder under load Mounted device falls: equipment damage; impact injury to patient/staff below Low–Med Med 4 perimeters; engineered chamfers, hole compensation and in-built supports; fit/tolerance QC before install; only listed devices mounted; weekly crack inspection; brim during printing for hook stability
R2 Print-quality defects (delamination, under-extrusion) Weakened part fails later in service Med (printer-dependent) Med Mandatory visual QC (reject stringing/misaligned layers); reference slicer profile and pre-sliced GCODE provided; light filament color makes defects visible
R3 Material aging — PETG embrittlement of latches/hooks Sudden fracture, device fall Med over time Med Explicit weekly flexibility check; storage away from sunlight/humidity; replace-don't-repair culture enabled by cheap local reprinting
R4 IPC hazard — contaminated mount surfaces in patient environment Cross-contamination / HAI contribution Med if uncleaned Med–High Approved agents (Surfanios, bleach 1:10, ethanol 70 %); smooth-surface QC; no embossed text; light colors reveal soiling; cleaning before first use; IPC-advisor approval gate
R5 Wrong fitment (tube/bottle diameter mismatch) Unstable mount, device fall Med if unverified Med Compatibility table with explicit diameters (19/25/32 mm tubes; 49/56 mm bottles); mandatory pre-installation dimensional validation
R6 Wall-mount fastener or substrate failure Mount detaches from wall with device Low Med Specified M5 / Ø4–5 mm fasteners; safety-lock feature in the wall-mount design; installation by technical staff
R7 Misuse beyond design intent (e.g. used as grab bar, patient support, or for heavier devices) Overload fracture; user fall or device fall Low–Med High Explicit out-of-scope statement (1.1); usage warnings (2.4); staff briefing at installation
R8 Sharp edges/burrs on printed parts Minor cuts; glove perforation Med pre-finishing Low Mandatory deburring in post-processing; visual QC
R9 Improper reprocessing (autoclaving) Deformed, weakened part returned to service unnoticed Low Med "Do not autoclave" is an explicit instruction; only surface disinfection permitted
R10 Unauthorized/undocumented deployment Use without biomed/IPC oversight Med in field settings Med Approval gate ("Approval required by: Biomed advisor, IPC advisor"); this documentation; packaging/storage instructions keep parts traceable to production

1.6.1 Adverse outcomes disclaimer

The Universal Mounting System V1 is open-source hardware in development, provided "as is", without warranty of any kind, under the MIT License. It is not a cleared, approved, or CE-marked device. Locally manufactured parts vary with printer, filament and operator; the design authors cannot verify any individual build. Foreseeable adverse outcomes of incorrect manufacture, installation, cleaning, or use include — without limitation — falling equipment causing injury to patients or staff, damage to medical devices, and infection-control breaches. Do not deploy in a clinical environment without documented approval from the responsible biomedical and IPC responsibles, completion of the QC procedures, and adherence to the maintenance schedule. Producers, installers and operating facilities assume full responsibility for verification and safe use of parts they manufacture and install.

1.6.2 Open issues

Live tracker: https://github.com/MSF3Dprinting/Universal-Mounting-System/issues

1.7 Assembly Manual and Maintenance Documentation

Build → install workflow:

  1. Print the required back and front parts (any FDM printer; PETG, light color; 0.2 mm layers; 4 perimeters; no added supports; brim recommended for hooks). Reference plate: UMS set.3mf / MK4S GCODE.
  2. Post-process: remove brim; clean stringing and surface imperfections; deburr sharp edges.
  3. QC: visual, dimensional and tolerance checks per Section 1.5; obtain biomed/IPC validation.
  4. Package/store: seal in zip-lock bags; store at room temperature away from direct sunlight and humidity.
  5. Install: mount the back part to the bed/tube/wall (B3 Hook-with-support: secure with a 6 mm zip tie; B4 Wall mount: M5 or Ø4–5 mm fasteners into a sound substrate), then slide in the appropriate front part.
  6. Clean before use with an approved agent (Section 2.3).

Maintenance documentation (required — multiple-use product):

  • Weekly in-service check after installation, focusing on: (a) cracks anywhere in the 3D-printed body, (b) surface cleanliness, (c) mechanical flexibility of latches and hooks — PETG can become brittle with age. Remove and replace any part that fails a check (local reprinting makes replacement the default corrective action).
  • Routine cleaning per IPC guidelines with the approved agents; never autoclave.
  • Front parts may be removed/swapped at any time for cleaning or reconfiguration without disturbing the installed back part — this is a deliberate maintainability feature of the two-part design.

2. Additional information

2.1 QMS Documentation

  • Documented QC procedure per part (visual, dimensional, tolerance, safety validation)
  • Defined approval gate (biomed + IPC advisors) and defined in-service check frequency (weekly) with named focus points.
  • Version history table and designer-of-record in the datasheet; source under Git version control.
  • Unit QC documentation log (printer, filament brand/batch, operator, QC result per part), sign the "approved by/tested by" fields per release, and tag repository releases used for production.

2.2 Hardware Legacy Plan

The project is structured for long-term reproducibility independent of the original maintainers:

  • License: MIT — permits use, modification, manufacture and redistribution, including commercially, in perpetuity.
  • Editable source, not just exports: full STEP models for every part (3D files/STEP/…) allow re-engineering in any CAD tool; STL/3MF/GCODE are provided for direct production.
  • Platform: public GitHub repository with version history; anyone may fork, so no single point of failure for availability.
  • Machine-agnostic manufacturing: "Any FDM 3D printer" with commodity PETG — no proprietary tooling, no sole-source components; the only purchased items are generic fasteners and zip ties.

2.3 Sterilization / Cleaning and Disinfection Plan

  • Approved reprocessing: surface cleaning/disinfection with Surfanios, bleach 1:10, or ethanol 70 %, following the relevant IPC guidelines of the facility.
  • Sterilization: not applicable and not permitted. Do not autoclave (thermal deformation of PETG). The device is intended as a non-sterile, non-patient-contact environmental accessory; the pending Spaulding classification
  • Design features supporting cleanability: smooth-surface QC requirement, no embossed text/symbols, light material color to reveal soiling, and tool-free removable front parts for off-mount cleaning.
  • Clean before first use and per facility schedule thereafter; cleanliness is part of the weekly inspection.

2.4 Usage Guidelines and Warnings

Guidelines

  1. Use only for the devices and mounting interfaces listed in the compatibility tables (1.4); verify fit before installation.
  2. Obtain biomed and IPC advisor approval before placing any part in a clinical area.
  3. Clean with approved agents before first use and per facility IPC schedule.
  4. Perform the weekly inspection (cracks, cleanliness, latch/hook flexibility); replace any suspect part immediately.
  5. Install wall mounts only into sound substrates with the specified fasteners; use the 6 mm zip tie with the Hook-with-support as instructed.
  6. Store spare parts sealed, at room temperature, away from direct sunlight and humidity.

Warnings

  • ⚠️ Not a patient-support device. Never use as a grab bar, handhold, restraint point, or to bear body weight.
  • ⚠️ Do not autoclave or expose to high heat — the part will deform and may fail without visible warning.
  • ⚠️ Do not mount devices other than those listed, or exceed the intended load; no quantitative load rating has been published yet.
  • ⚠️ Do not add engraved/embossed markings to parts; do not use dark filaments that hide soiling and defects.
  • ⚠️ Do not use any part with cracks, whitened stress marks, or stiff/brittle latches — replace it.
  • ⚠️ Falling equipment hazard: position mounts so that a failure would not drop a device onto a patient, infant cot, or walkway where avoidable.

About

Universal Mounting System V1 (UMS V1) is set of mutually compatible attachments, hooks, holders and clamps used to mount various devices and accessories on the wall, beds and other objects. The UMS allows easy attachment, removal and changing of the devices.

Resources

Readme
MIT license
Activity
Custom properties

Stars

0 stars

Watchers

0 watching

Forks

0 forks
Report repository

Releases

Packages

Contributors

You are visitor 01402792 — GeezerVison does not store the pages you look up.
Best viewed with Netscape Navigator 4.7 or Internet Explorer 6 at 800×600. No cookies required.