GV

GeezerVison

The World Wide Web, restored to 800 pixels of glory.

Home|About|Gaming cabinet|Archives|How it works|Starter sites

On this site

  • Sign in
  • ActionsCI/CD and automation
  • CodespacesCloud development environments
  • All Features
  • Documentation
  • Training
  • Certifications
  • Build with Copilot
  • Customize Copilot
  • Integrations
  • Trending repos
  • GitHub Sponsors
  • Developer articles
  • Enterprises

This view

  • Original page
  • New address

GeezerVison is showing github.com — GitHub - sahgyan9/Cryocon_22C_Temperature_Controller_Python-File: Python RS-232 driver, Tkinter GUI, and PID overshoot study for a Cryo-con 22C controller on a Janis ST-LN-500 cryogenic probe station.

BranchesTags
Open more actions menu

Latest commit

 

History

31 Commits
31 Commits

Folders and files

NameName
Last commit message
Last commit date
Cryocon-22C_and_Wayne_Kerr
Cryocon-22C_and_Wayne_Kerr
 
 
archive
archive
 
 
data
data
 
 
data_backup_20260904
data_backup_20260904
 
 
docs
docs
 
 
figures
figures
 
 
labview
labview
 
 
.gitignore
.gitignore
 
 
AGENTS.md
AGENTS.md
 
 
Cryocon 22C Overshoot Study.pdf
Cryocon 22C Overshoot Study.pdf
 
 
FINAL_REPORT.html
FINAL_REPORT.html
 
 
FINAL_REPORT.md
FINAL_REPORT.md
 
 
FROSTBYTE_PLAN.md
FROSTBYTE_PLAN.md
 
 
LABVIEW_RAMP_CONTROL_GUIDE.md
LABVIEW_RAMP_CONTROL_GUIDE.md
 
 
OVERSHOOT_TUNING_LOG.md
OVERSHOOT_TUNING_LOG.md
 
 
PMN-0.3PT_Dielectric_Spectroscopy_Technical_Report.pdf
PMN-0.3PT_Dielectric_Spectroscopy_Technical_Report.pdf
 
 
README.md
README.md
 
 
cryocon_controller.py
cryocon_controller.py
 
 
cryocon_gui.py
cryocon_gui.py
 
 
live_view.py
live_view.py
 
 
make_final_figures.py
make_final_figures.py
 
 
monitor_only.py
monitor_only.py
 
 
pid_controller.py
pid_controller.py
 
 
read_full_table.py
read_full_table.py
 
 
requirements.txt
requirements.txt
 
 
run_gui.bat
run_gui.bat
 
 
staged_ramp_test.py
staged_ramp_test.py
 
 
write_pid_table02.py
write_pid_table02.py
 
 

Repository files navigation

Cryocon 22C — Janis ST-LN-500 Probe Station Control

Temperature control for a Janis Research ST-LN-500_1-4CX LN₂ cryogenic probe station via a Cryo-con 22C controller on COM5.

Start here: FINAL_REPORT.html — the overshoot study, the settings that fixed it, and the LabVIEW SCPI sequence. Open it in a browser; it is self-contained and prints cleanly.

Current tuning (validated 2026-09-05)

LOOP 1:PGAIN   40        proportional gain
LOOP 1:IGAIN   900       integral time, SECONDS (not a gain)
LOOP 1:DGAIN   0         derivative off
LOOP 1:RANGE   HI        50 W full scale
LOOP 1:MAXPWR  70        percent
LOOP 1:RATE    1.0       K/min (2.0 works equally well and is twice as fast)
LOOP 1:TYPE    RAMPP

Peak overshoot +0.22 K, down from +4.08 K. PID Table 2 in the instrument's memory now carries these Zone 1 values as well, so RAMPT/TABLE mode is safe above 300 K.

Ordering rule that matters

Send CONTROL before the target setpoint, and send the setpoint last — writing the setpoint is what arms the ramp. Setting a target while control is OFF makes the heater slam to its power limit the moment control is switched on.

Serial Interface & Communication Speed

  • Port: COM5 (or assigned USB-serial COM port)
  • Baud Rate: 57,600 (8 Data Bits, 1 Stop Bit, No Parity, \r\n Line Termination)
  • Hardware Upgrade Note: Upgraded from the factory 9,600 baud to 57,600 baud on the instrument front panel (System -> Remote -> RS-232 Rate: 57600).
  • Telemetry Throughput & ADC Limits:
    • The Cryo-con 22C has a fixed 24-bit delta-sigma ADC running at 10 Hz (100 ms).
    • At 57,600 baud, single SCPI query roundtrips take ~25–30 ms (down from ~120 ms at 9600).
    • The Python GUI (cryocon_gui.py) features an Ultra (0.10 s / 10 Hz) polling option that streams stage temperature, heater output, and setpoints at the full hardware ADC limit with zero lag.

Layout

Path Contents
FINAL_REPORT.html / .md The report
OVERSHOOT_TUNING_LOG.md Full technical log — every attempt, superseded theories, safety config
data/ 26 measured runs + the PID table backup taken before overwriting
figures/ Report figures (regenerate with make_final_figures.py)
docs/ Cryocon manual, project objective, dev notes
labview/ LabVIEW port — build spec, 22C simulator, ActiveX test harness
archive/ Superseded scripts and the earlier report's generators
data_backup_20260904/ Independent backup taken mid-study
Cryocon-22C_and_Wayne_Kerr/ Unified suite: Wayne Kerr 6510B Impedance Analyzer + Cryo-con 22C

Integrated Suite: Wayne Kerr 6510B & Cryo-con 22C (PMN-0.3PT Characterization)

The folder Cryocon-22C_and_Wayne_Kerr/ contains the unified impedance spectroscopy software used for characterizing PMN-0.3PT ($0.70\text{Pb}(\text{Mg}{1/3}\text{Nb}{2/3})\text{O}_3 - 0.30\text{PbTiO}_3$):

  • unified_gui.py: Unified GUI orchestrating simultaneous temperature ramps and multi-frequency dielectric impedance sweeps with live $\varepsilon'(T)$ visualization.
  • Latest Data: Subfolder Cryocon-22C_and_Wayne_Kerr/Data/ contains the latest 57-point dataset (300 K to 412 K) for PMN-0.3PT.
  • Launcher: Run Cryocon-22C_and_Wayne_Kerr\run_gui.bat or python Cryocon-22C_and_Wayne_Kerr/unified_gui.py.
  • Documentation: See Cryocon-22C_and_Wayne_Kerr/README.md and LEARNING_LOG.md for full physical and operational details.

Scripts

Script Purpose
staged_ramp_test.py Ramp runner for tests. Sends STOP on every exit path — use this for anything unattended.
live_view.py Live temperature/heater display. Tails the log file, so it does not occupy COM5 while a test runs.
write_pid_table02.py Writes PID Table 2 to instrument NVRAM. Backs nothing up itself — take a copy first.
make_final_figures.py Regenerates every report figure from data/.
read_full_table.py Dumps the PID tables currently in the instrument.
monitor_only.py Read-only monitor.
cryocon_gui.py GUI (run_gui.bat).

Only one process can hold COM5. Close the GUI and stop any running test before starting another.

Open items

  • 77–300 K zone is untuned. Its table values are likely wrong for the same units reason as Zone 1, but LN₂ changes the dynamics and there is no data yet. LN₂ is on order — retune once it arrives.
  • Over Temperature Disconnect (Channel A, 470 K) is set by hand on the front panel and is not remotely settable or readable. Re-enter it manually if the instrument is ever reset. (Updated from 460 K).

About

Python RS-232 driver, Tkinter GUI, and PID overshoot study for a Cryo-con 22C controller on a Janis ST-LN-500 cryogenic probe station.

Topics

cryogenicsexperimental-physicsinstrument-controlpid-controlpyserialpythonscpi

Resources

Readme
Activity

Stars

0 stars

Watchers

0 watching

Forks

0 forks
Report repository

Releases

Packages

Contributors

Languages

You are visitor 01402775 — GeezerVison does not store the pages you look up.
Best viewed with Netscape Navigator 4.7 or Internet Explorer 6 at 800×600. No cookies required.